| Model Number |
E163 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Codes |
Shock from Patient Lead(s) (3162); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/30/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited noise that was oversensed during what appeared to be a procedure.This icd remains in service.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited noise that was oversensed during what appeared to be a procedure.This icd remains in service.No adverse patient effects were reported.Additional information was received and this icd delivered inappropriate anti-tachycardia pacing (atp) and four shocks due to a magnet not being placed during a procedure.This icd remains in service.No additional adverse patient effects were reported.
|
| |
|