• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN AG JUAREZ (MMJ) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN AG JUAREZ (MMJ) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 3.5NCF
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that even when air was inserted into the tracheostomy tube's cuff during the pre-cuff check, it would deflate.There was no patient involvement.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3 and h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that even when air was inserted into the tracheostomy tube's cuff during the pre-cuff check, it would deflate.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
COVIDIEN AG JUAREZ (MMJ)
1181 hennequen ave.
juarez,chi 32575
MX  32575
Manufacturer (Section G)
COVIDIEN AG JUAREZ (MMJ)
1181 hennequen ave.
juarez,chi 32575
MX   32575
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key21961044
Report Number9681384-2025-00487
Device Sequence Number13074657
Product Code JOH
UDI-Device Identifier20884521779553
UDI-Public20884521779553
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K182861
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3.5NCF
Device Catalogue Number3.5NCF
Device Lot Number24F0936JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/30/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/23/2025
Supplement Date Received by Manufacturer08/22/2025
Initial Report FDA Received Date05/05/2025
Supplement Report FDA Received Date08/25/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/04/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-