| Model Number |
A20710A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported the grasping forceps had something break on the grasping piece.The issue occurred during a therapeutic bladder stone removal procedure.There was a delay of less than 30 minutes under sedation, with device inside the patient.The procedure was completed using an olympus back-up device.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
The mdr supplemental report is being submitted to provide updated fields and final investigation results.Updated fields: d9, h2, h3, h4, h6, h11 the device was returned for evaluation and the customer's allegation was confirmed.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: cause traced to component failure.Olympus will continue to monitor the performance of this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from the customer.
|
| |
|