ZIMMER BIOMET, INC. AVENIR COMPLETE STEM CEMENTLESS COLLARED HIGH OFFSET 12/14 TAPER SIZE 3; PROSTHESIS, HIP Back to Search Results
Catalog Number 574202030
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Periprosthetic fracture (4902)
Date of Event 04/25/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Proposed component code: mechanical (g04), stem.G2: foreign: australia.The customer has indicated that the product will not be returned to zimmer biomet for investigation as it was not returned by the hospital.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient underwent an initial hip procedure.Subsequently, the patient was revised approximately 3 weeks later due to periprosthetic fracture.Attempts have been made, and no further information has been provided.
 
Event or Problem Description
No further information at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were {update/corrected}.Updated: d4; d5; g3; h2; h3; h4; h6.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record identified no deviations or anomalies during manufacturing.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: the left hip arthroplasty components are anatomically aligned.No periprosthetic fracture or other abnormality is identified.A definitive root cause cannot be determined.Unable to confirm complaint.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVENIR COMPLETE STEM CEMENTLESS COLLARED HIGH OFFSET 12/14 TAPER SIZE 3
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key21961667
Report Number0001822565-2025-01250
Device Sequence Number17960267
Product Code LZO
UDI-Device Identifier00889024509153
UDI-Public(01)00889024509153(17)291202(10)3216178
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K182048
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number574202030
Device Lot Number3216178
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2025
Supplement Date Received by Manufacturer06/23/2025
Initial Report FDA Received Date05/05/2025
Supplement Report FDA Received Date06/27/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/02/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age82 YR
Patient SexFemale
-
-