DEPUY IRELAND - 3015516266 ATTUNE REV IMPLNT ASSEM WRENCH; WRENCHES Back to Search Results
Catalog Number 254600420
Medical Device Problem Codes Calibration Problem (2890); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/14/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
 
Event or Problem Description
It was reported that the assembly wrench had a screw back out of the torque plate and will not properly gauge the torque being put on the wrench.All pieces were retrieved and the patient was not harmed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: an analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product or product id numbers are received at a later date, the investigation will be updated as applicable.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d4 (lot), d9, h4.Corrected: d4 (primary udi #).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: it was reported that the assembly wrench had a screw back out of the torque plate and will not properly gauge the torque being put on the wrench.All pieces were retrieved and the patient was not harmed.The product was returned to depuy synthes for evaluation.Visual analysis of the returned sample revealed that the calibration plate and the screws were missing.This condition may occur during the cleaning/sterilization process.As these procedures are often repeated, there is an increased risk of components going missing.However, without concrete evidence or further information, it is not possible to determine a root cause.A dimensional inspection was not performed since it was not applicable to the complaint condition.A functional test was unable to performed due to the absence of essential components, however with this observed condition the assembly issues can be confirmed.The overall complaint was confirmed as the observed condition of the attune rev implnt assem wrench would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Corrected: h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE REV IMPLNT ASSEM WRENCH
Common Device Name
WRENCHES
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key21962273
Report Number1818910-2025-07062
Device Sequence Number18438775
Product Code HXC
UDI-Device Identifier10603295440338
UDI-Public(01)10603295440338(10)GB105008
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number254600420
Device Lot NumberGB105008
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/14/2025
Supplement Date Received by Manufacturer05/23/2025
09/16/2025
01/12/2026
Initial Report FDA Received Date05/05/2025
Supplement Report FDA Received Date05/26/2025
09/17/2025
01/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/06/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age72 YR
Patient SexFemale
-
-