Model Number N/A
Medical Device Problem Code Material Frayed (1262)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/07/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by the customer guidewire came unraveled while inserting into the patient.An additional guidewire was needed for access causing a delay in therapy.Complainant states there was no adverse impact on the patient.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and it will be evaluated.A supplemental will be submitted with evaluation results.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a broken guidewire was confirmed.The product returned for evaluation was one guidewire.Additionally, one photograph was also submitted by the complainant for review.No additional information was observed in the photographs which changed the conclusion of the investigation.The investigation findings are consistent with damage caused by retraction of the guidewire against the bevel of the introducer needle.The returned product sample was evaluated and the guidewire was confirmed to be broken which allowed the outer coil wire to become unraveled.Microscopic examination of the fracture sites revealed the following: narrowing of the wire cross-section near the fracture site, which is a characteristic feature of a strong pull on the wire.The weld tip of the wire was missing and could not be accounted for during the evaluation.Normal movement of the guidewire is away from the sharpened bevel and will not damage the wire.If the guidewire direction is reversed, the guidewire is then pulled against the sharpened edge of the needle bevel and can cause shearing damage of the wire and/or cause the wire to become stuck within the needle.An examination of the wire structure revealed no potential damage/defect related to manufacture of the product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROVENA MIDLINE
Common Device Name
MIDLINE CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kayla bevins
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key21962447
Report Number3006260740-2025-02826
Device Sequence Number15260901
Product Code PND
UDI-Device Identifier00801741154126
UDI-Public(01)00801741154126(11)241126(17)260228(10)REJW2527
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K162900
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Model NumberN/A
Device Catalogue NumberS4254108DG
Device Lot NumberREJW2527
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/14/2025
Supplement Date Received by Manufacturer06/25/2025
Initial Report FDA Received Date05/05/2025
Supplement Report FDA Received Date07/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/26/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-