| Model Number |
WA02944A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported the subject device exhibited broke brush off in the side channel.The issue occurred during reprocessing.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d9, h2, h3, h4, h6, h11.The device was returned to olympus for inspection, and the reported failure was confirmed.A root cause could not be identified.Based on the results of the investigation, the most probable cause for the malfunction is traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|