• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. AERIS EVO; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. AERIS EVO; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number VT2110X24B
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/04/2024
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging a pressure sensor mismatch error code occurred after updating an aeris evo/trilogy evo ventilator firmware 1.6.10.00.There was no harm or injury reported.There was no reported patient involvement.It is noted that this was an out of box failure and that sometimes after a few hours of run-in, the error clears, and then sometimes it remains.The internal machine flow sensor needs to be replaced to restore the aeris evo/trilogy evo device to working order per the recommendation of vitalaire italy, medical equipment provider.The device has not been returned to the manufacturer and therefore, unable to confirm the alleged malfunction.At this time a final report will be submitted stating the recommended resolution.If further information becomes available at a later date, a supplemental report will be submitted.No further investigation is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AERIS EVO
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key21964874
Report Number2518422-2025-105217
Device Sequence Number18251964
Product Code CBK
UDI-Device Identifier00606959062962
UDI-Public(01)00606959062962
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberVT2110X24B
Device Catalogue NumberVT2110X24B
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/04/2024
Initial Report FDA Received Date05/05/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/27/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-