| Catalog Number |
SRF-695 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
03/27/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported "rupture".This record is for the right side.The device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the product analysis performed, the assessments of the complaint codes are: - rupture: observed broken device assessed as fold flaw opening and observed a missing piece of shell assessed as inconclusive.As per the investigation procedure, creases, wear abrasion and non penetrating nicks were observed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.Additional, changed, and/or corrected data: d.9., h.2., h.3., h.6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "rupture".This record is for the right side.The device was explanted.
|
| |
|