• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TPRLC 133 MP TYPE1 PPS HO 18.0; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TPRLC 133 MP TYPE1 PPS HO 18.0; PROSTHESIS, HIP Back to Search Results
Catalog Number 51-107180
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/10/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The product has been received by zimmer biomet.The investigation is currently in process.Once the investigation has been completed, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that there was a hole in the sterile packaging.The initial implant didn¿t make it to the sterile field.It was noticed just after opening.The procedure was completed using another device.There is no further information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4;g3;h2;h3;h6 product returned and evaluated.Visual evaluation of the returned product/provided photos identified damage to the sterile packaging (pouch).Sterility had been compromised.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The reported event was not related to a combination of products; therefore, a compatibility review is not applicable.Medical records were not provided.The condition of the device when it left zimmer biomet is considered conforming to specification.The root cause of the reported event can be attributed to transit damage and a packaging design issue.The reported event has been confirmed by evaluation of the returned product and provided photos.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 MP TYPE1 PPS HO 18.0
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key21967059
Report Number0001825034-2025-01328
Device Sequence Number18246365
Product Code KWA
UDI-Device Identifier00880304513594
UDI-Public(01)00880304513594(17)290618(10)6554884
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K110400
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number51-107180
Device Lot Number6554884
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/16/2025
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/10/2025
Supplement Date Received by Manufacturer08/13/2025
Initial Report FDA Received Date05/06/2025
Supplement Report FDA Received Date08/14/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/18/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-