BOLTON MEDICAL, INC. TREO ABDOMINAL STENT-GRAFT SYSTEM; STENT, ENDOVASCULAR GRAFT, AORTIC Back to Search Results
Catalog Number 28-L2-15-120U
Medical Device Problem Code Device Endoleak (4074)
Health Effect - Clinical Codes Aneurysm expansion (4597); Clinical Endoleak (4843)
Type of Reportable Event Serious Injury
Event or Problem Description
"the doctor called me about a previous evar which had a leak.After images were acquired, it was then realized the device which was implanted was a treo.Scan shows that the aorta has remodeled.The renal to bifurcation length is significantly longer than pre evar scan.The fabric line appear to be roughly 7mm below the lowest renal (right renal), disease seems to have dilated the neck, and there is a type 1a leak.There also is an apparent type 2 endoleak.The right iliac limb seems to have sucked up into the aneurysm and very little is left in the right common iliac artery.There is potential of a type 1b leak." patient outcome: "patient will likely be referred to another institute for a pmeg repair".
 
Additional Manufacturer Narrative
This complaint was involved with three devices.Device 1 is being reported under mdr-2247858-2025-00115, device 2 is being reported under mdr-2247858-2025-00116, and device 3 is being reported.Bolton medical, inc.(d/b/a terumo aortic), herein known as the "company", is submitting this report pursuant to 21 cfr part 803, has made reasonable efforts to obtain complete information, and has provided as much as is available to the company as of the submission date of this report.This report is based on information obtained by the company, which may not have been able to fully investigate or verify prior to the date the report was required by the regulations.Any required fields that are unpopulated are blank because the information is currently unknown, unavailable, or not applicable.This report does not constitute an admission or a conclusion by anyone that the device caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the certain regulations, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
"the doctor called me about a previous evar which had a leak.After images were acquired, it was then realized the device which was implanted was a treo.Scan shows that the aorta has remodeled.The renal to bifurcation length is significantly longer than pre evar scan.The fabric line appear to be roughly 7mm below the lowest renal (right renal), disease seems to have dilated the neck, and there is a type 1a leak.There also is an apparent type 2 endoleak.The right iliac limb seems to have sucked up into the aneurysm and very little is left in the right common iliac artery.There is potential of a type 1b leak." patient outcome: "patient will likely be referred to another institute for a pmeg repair".
 
Additional Manufacturer Narrative
This complaint was involved with three devices.Device 1 is being reported under mdr-2247858-2025-00115, device 2 is being reported under mdr-2247858-2025-00116, and device 3 is being reported under mdr-2247858-2025-00117.Bolton medical, inc.(d/b/a terumo aortic), herein known as the "company", is submitting this report pursuant to 21 cfr part 803, has made reasonable efforts to obtain complete information, and has provided as much as is available to the company as of the submission date of this report.This report is based on information obtained by the company, which may not have been able to fully investigate or verify prior to the date the report was required by the regulations.Any required fields that are unpopulated are blank because the information is currently unknown, unavailable, or not applicable.This report does not constitute an admission or a conclusion by anyone that the device caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the certain regulations, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TREO ABDOMINAL STENT-GRAFT SYSTEM
Common Device Name
STENT, ENDOVASCULAR GRAFT, AORTIC
Manufacturer (Section D)
BOLTON MEDICAL, INC.
799 international parkway
sunrise FL 33325
Manufacturer Contact
manuel benitez
799 international parkway
sunrise, FL 33325
9548389699
MDR Report Key21967607
Report Number2247858-2025-00117
Device Sequence Number8112330
Product Code MIH
UDI-Device Identifier08435761052142
UDI-Public08435761052142
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P190015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date06/03/2024
Device Catalogue Number28-L2-15-120U
Device Lot Number2105220202
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/11/2025
Supplement Date Received by Manufacturer04/11/2025
Initial Report FDA Received Date05/06/2025
Supplement Report FDA Received Date08/07/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age81 YR
Patient SexMale
-
-