• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NORIS MEDICAL LTD TUFF UNICON IMPLANT; TUFF UNICON DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NORIS MEDICAL LTD TUFF UNICON IMPLANT; TUFF UNICON DENTAL IMPLANT Back to Search Results
Model Number NMCF4211
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/09/2025
Type of Reportable Event Serious Injury
Event or Problem Description
A healthcare professional reported that surgeon unable to get appropriate torque on implant site #4, nmcf4211, lot#9001322 implant lacked primary stability.The implant was removed during implant placement with no report of observable clinical symptoms.According to the information, the patient has hypertension.The device was forwarded to the manufacturer.No other medical issues were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUFF UNICON IMPLANT
Common Device Name
TUFF UNICON DENTAL IMPLANT
Manufacturer (Section D)
NORIS MEDICAL LTD
hataasia 8
nesher, 36888 08
IS  3688808
MDR Report Key21967647
Report Number3012141159-2025-00288
Device Sequence Number8131269
Product Code DZE
UDI-Device Identifier07290116153421
UDI-Public07290116153421
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 05/06/2025
Report Date (Section F) 05/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberNMCF4211
Device Catalogue NumberNMCF4211
Device Lot Number9001322
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date04/22/2025
Device Age7 MO
Event Location Outpatient Treatment Facility
Date Report to Manufacturer05/06/2025
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date05/06/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age56 YR
Patient SexMale
Patient Weight88 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
Date Report Sent to FDA05/06/2025
-
-