| Model Number |
INGENIA ELITION S |
| Medical Device Problem Code |
Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
Hematoma (1884)
|
| Date of Event |
04/12/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Philips received a report about an attraction incident on an elition mri in which a hospital employee was injured.
|
| |
|
Additional Manufacturer Narrative
|
|
No functional tests were performed as this issue was caused by a use error by bringing a non-mr safe bed into the examination room.Based on the provided information there is no indication of a malfunction of the mri system.It is a known issue that no magnetic materials should be brought into the examination room.Based on the information available, the cause of the reported problem was a user error.No information was made available related to the outcome for the user other then that an injury occurred.
|
| |
|
Search Alerts/Recalls
|