• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION MOSAIC AORTIC BIOPROSTHETIC HEART VALVE; REPLACEMENT HEART-VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION MOSAIC AORTIC BIOPROSTHETIC HEART VALVE; REPLACEMENT HEART-VALVE Back to Search Results
Model Number 305U227
Medical Device Problem Codes Gradient Increase (1270); Patient Device Interaction Problem (4001); Device Stenosis (4066)
Health Effect - Clinical Codes Aortic Valve Stenosis (1717); Dyspnea (1816)
Date of Event 04/08/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Medtronic received information that 14 years and 8 months post implant of this aortic bioprosthetic valve, the patient is being evaluated for a transcatheter valve-in-valve replacement.The reason for evaluation was reported as severe aortic stenosis, a mean gradient of 32mmhg, a peak systolic velocity (psv) of 3.6m/s, a left ventricular ejection fraction (lvef) of 15-20%, an aortic valve area of 0.57cm2 and a stroke volume of 42.It was also reported that the patient had symptoms of shortness of breath.No intervention or adverse patient effects were reported.Subsequently, 7 days later, a transcatheter valve was implanted valve-in-valve.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOSAIC AORTIC BIOPROSTHETIC HEART VALVE
Common Device Name
REPLACEMENT HEART-VALVE
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key21968836
Report Number2025587-2025-03430
Device Sequence Number8260533
Product Code DYE
UDI-Device Identifier00613994479228
UDI-Public00613994479228
Combination Product (Y/N)N
Initial Reporter StateMT
Initial Reporter CountryUS
PMA/510(K) Number
P990064
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/11/2015
Device Model Number305U227
Device Catalogue Number305U227
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/08/2025
Initial Report FDA Received Date05/06/2025
Date Device Manufactured06/16/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age73 YR
Patient SexMale
Patient Weight78 KG
-
-