| Model Number |
VP-902-X |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Unintended Radiation Exposure (4565); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/21/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, during the bypass surgery, when on vascular anastomosis, the needle broke on the swage/central part.Imaging test was then performed.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: vp-902-x surgipro*ii 8-0 60cm mv1355da, vp-902-x, lot# d4c4401y.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b1, b2 (removed intervention req), b5, g3, h6.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during the bypass surgery, when on vascular anastomosis, the needles broke on the swage/central part.A new product was used to resolve the issue.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: h1.New information has been received pertaining to the event.This event has been reassessed and the reportability has been determined to be a malfunction reportable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d4 (udi), d9, g3, h3, h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.Two devices were available for evaluation.Visual inspection noted two sutures and two needles.The needles were received broken in the middle.Upon microscopic inspection of the needles, normal crimp and swage marks were noted.Tool marks were noted near the break sites.It was reported that the needles had broken.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|