| Model Number |
E161 |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/21/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) displayed an error message and when examined, it had entered safety mode.Additionally, the device was beeping as a result.The device was explanted and a new device was implanted.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Interrogation of the device confirmed it was operating in safety mode and that critical therapy remained available.Review of device memory identified an error.The error resulted in software resets performed in an attempt to correct an identified memory inconsistency.The corrupted memory was corrected in the laboratory and the device reverted to normal operation.The cause of the memory inconsistency was not able to be determined through laboratory testing but was likely the result of exposure to radiation, either therapeutic or environmental.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) displayed an error message and when examined, it had entered safety mode.Additionally, the device was beeping as a result.The device was explanted and a new device was implanted.No additional adverse patient effects were reported.The device has been returned and analyzed.
|
| |
|