| Model Number |
HYBRID FIBERTAK WITH NEEDLES, KNEE |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 04/08/2025, a sales representative reported via email that (qty.2) of an ar-3770sp knee fibertak hybrid (one knotless and one knotted) failed on insertion and pulled out multiple times.The case was completed using two ar-3770sp knee fibertak hybrids from lot 15339225 without issues.This was discovered during a procedure on (b)(6) 2025, with no reported patient harm.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
Additional information was provided: this issue was identified during an mpfl procedure.All affected components were removed from the patient before proceeding with the new ar-3770sp knee fibertak hybrids.There was no reported case delay.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to improper bone prep and/or shallow driver engagement depth.
|
| |
|