This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: the complaint device was discarded by the customer, therefore not returned to j&j medtech orthopaedics and unavailable for a physical evaluation.A manufacturing record evaluation was performed for the finished device, and no non-conformances related to the reported complaint condition were identified.Based on the current available information, we cannot determine a root cause for the reported failure.If additional information is received in the future, we will reopen the complaint and perform the investigation as appropriate.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
|
It was reported that during an arthroscopic shoulder rotator cuff tear surgical procedure for a rotator cuff tear on the shoulder, while using the 4.5hlx adv br anc-dyna 3-sut device a bone hole was created with an awl as usual and the anchor was inserted, but part way through the surgeon noticed an abnormal noise and when he checked the anchor it was cracked and the anchor part was broken.He had the anchor removed in the reverse direction and all of the anchor was retrieved.The anchor was successfully inserted and sutured using as spare device.There was a thirty minute delay to the procedure.There were no reports of injuries, medical intervention or prolonged hospitalization.No additional information was provided.
|