• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BTG ROUND LAKE VISUAL-ICE CRYOABLATION SYSTEM; UNIT, CRYOSURGICAL, ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BTG ROUND LAKE VISUAL-ICE CRYOABLATION SYSTEM; UNIT, CRYOSURGICAL, ACCESSORIES Back to Search Results
Model Number FPRCH6000
Medical Device Problem Codes Insufficient Cooling (1130); Display or Visual Feedback Problem (1184); Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/09/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported insufficient ice occurred.A visual-ice cryoablation system was chosen to complete a cryoablation procedure.A channel issue transpired on the all the top channels resulting in insufficient ice being formed by the system.The gas reading on the machine was not the same as the cylinder even when the cylinder was full.The original system was used to complete the procedure.No patient complications occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISUAL-ICE CRYOABLATION SYSTEM
Common Device Name
UNIT, CRYOSURGICAL, ACCESSORIES
Manufacturer (Section D)
BTG ROUND LAKE
4364 round lake road
arden hills MN 55112
Manufacturer (Section G)
GALIL MEDICAL
4364 round lake road
arden hills MN 55112
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key21971212
Report Number2124215-2025-29017
Device Sequence Number8692015
Product Code GEH
UDI-Device Identifier00867379000006
UDI-Public(01)00867379000006(11)170830(10)VL0317
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K113860
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberFPRCH6000
Device Catalogue NumberFPRCH6000
Device Lot NumberVL0317
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/09/2025
Initial Report FDA Received Date05/06/2025
Date Device Manufactured08/30/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-