AMO MANUFACTURING NETHERLANDS TECNIS SIMPLICITY; LENS, MULTIFOCAL INTRAOCULAR Back to Search Results
Model Number DRN00V
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Eye / Vision Problem (4471)
Date of Event 03/17/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H3: the device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of records including device history and complaint trending will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the preloaded intraocular lenses (iol) were explanted from patient's left and right eye as the patient was unable neuroadapt to the lenses.The lenses were explanted and replaced with a non jnn lenses in a secondary procedure.There was no patient injury.There was no medical/surgical intervention required.No other information is available.
 
Additional Manufacturer Narrative
Additional information: section d9: device available for evaluation: yes.Section d9: returned to manufacturer on: 5/6/2025.Section h3: device evaluated by manufacturer: yes.Device evaluation: visual inspection under magnification was performed on the suspect product and found the lens was cut in half.No issues were observed on either half of the lens.No further evaluation was performed.Conclusion: no issues that could cause or contribute to the complaint issues reported were identified during product evaluation.The ¿lens cut¿ observed during the product evaluation could not be confirmed to be related to the manufacturing or design process.Based on the complaint investigation results, the product was released within specifications.No product deficiency or product malfunction could be identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SIMPLICITY
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO MANUFACTURING NETHERLANDS
van swietenlaan 5
groningen, groningen 9728 NX
NL  9728 NX
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key21971798
Report Number3012236936-2025-000126
Device Sequence Number20898122
Product Code MFK
UDI-Device Identifier05050474810778
UDI-Public(01)05050474810778(17)270528
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberDRN00V
Device Catalogue NumberDRN00V0205
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/08/2025
Supplement Date Received by Manufacturer05/28/2025
Initial Report FDA Received Date05/06/2025
Supplement Report FDA Received Date06/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/28/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age41 YR
Patient SexFemale
-
-