MAUDE Adverse Event Report: INSPIRE MEDICAL SYSTEMS INSPIRE MODEL 4340 RESPIRATORY SENSING LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INSPIRE MEDICAL SYSTEMS INSPIRE MODEL 4340 RESPIRATORY SENSING LEAD Back to Search Results
Model Number 4340
Medical Device Problem Code Device Sensing Problem (2917)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/24/2025
Type of Reportable Event Malfunction
Event or Problem Description
During a routine examination, the physician identified a device sensing issue.Imaging was performed, and the distal tip of the sensing lead was suspected to have separated from the lead body.The patient was brought in for revision surgery, which confirmed that the distal sensor tip had detached and migrated into the pleural.Due to the potential risk involved in retrieval, the physician elected to abandon the migrated sensor tip.A new sensing lead was successfully implanted, restoring therapy.The procedure was completed without further complications, and the issue has been resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRE MODEL 4340 RESPIRATORY SENSING LEAD
Common Device Name
INSPIRE MODEL 4340 RESPIRATORY SENSING LEAD
Manufacturer (Section D)
INSPIRE MEDICAL SYSTEMS
5500 wayzatablvd.
suite 1600
golden valley MN 55416
Manufacturer (Section G)
INSPIRE MEDICAL SYSTEMS
5500 wayzata blvd.
suite 1600
golden valley MN 55416
Manufacturer Contact
vang yang
5500 wayzata blvd.
suite 1600
golden valley, MN 55416
7632057970
MDR Report Key21973106
Report Number3007666314-2025-00269
Device Sequence Number15272108
Product Code MNQ
UDI-Device Identifier10855728005632
UDI-Public10855728005632
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P130008
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4340
Device Catalogue Number900-009-002
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2025
Initial Report FDA Received Date05/06/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-