• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH CONCERTO; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH CONCERTO; COCHLEAR IMPLANT Back to Search Results
Model Number MI1000 MED-EL CONCERT
Medical Device Problem Code Impedance Problem (2950)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/22/2014
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event or Problem Description
The audiologist asked an integrity test for the user to clarify if there is a failure in the left implant.For several years, in situ measurements showed a progressive number of channels with high impedance.A ct scan will be scheduled.Re-implantation is being considered.
 
Event or Problem Description
The audiologist asked for an integrity test for the user to clarify if there is a failure in the left implant.A ct scan will be scheduled.Re-implantation is being considered.
 
Additional Manufacturer Narrative
Additional information: according to the currently available information, damage to the active electrode likely caused by minute device mobility is suspected.However, to determine an exact root cause a device investigation of the explanted device is necessary.Re-implantation is considered but no date has been scheduled yet.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONCERTO
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key21974511
Report Number9710014-2025-00408
Device Sequence Number12581072
Product Code MCM
UDI-Device Identifier09008737062682
UDI-Public(01)09008737062682
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/01/2014
Device Model NumberMI1000 MED-EL CONCERT
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/02/2025
Supplement Date Received by Manufacturer05/02/2025
Initial Report FDA Received Date05/07/2025
Supplement Report FDA Received Date07/01/2025
Date Device Manufactured04/10/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age3 YR
Patient SexFemale
-
-