• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) CARESTATION 750C; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) CARESTATION 750C; ANESTHESIA GAS MACHINE Back to Search Results
Model Number A1
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/09/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Ge healthcare (gehc) reported a field modification for an issue that prevents effective mechanical ventilation in volume control ventilation (vcv) mode per 21 cfr 806 on 02-apr-2025.The fda recall numbers are: z-1624-2025, z-1625-2025, z-1626-2025, z-1627-2025, z-1628-2025, z-1629-2025, z-1630-2025, z-1631-2025, z-1632-2025, z-1633-2025, z-1634-2025, z-1635-2025, z-1636-2025, z-1637-2025, z-1638-2025.Customers were sent a letter explaining the issue and providing instructions for inspecting for effective ventilation in volume control ventilation (vcv) mode.Gehc will correct all devices that fail the ventilation screening test at no cost.Block a: no patient involvement.Legal manufacturer: hcs madison - 3030 ohmeda dr, usa madison, wi 53718.
 
Event or Problem Description
As a result of an inspection that was completed as part of the correction initiated by ge healthcare (gehc) on 02-apr-2025, (gehc field action number 34142) this unit was identified as having an issue that prevents effective mechanical ventilation in volume control ventilation (vcv) mode.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARESTATION 750C
Common Device Name
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi 21402 8
CH  214028
Manufacturer (Section G)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi 21402 8
CH   214028
Manufacturer Contact
anthony amenson
3030 ohmeda drive
madison, WI 53718
MDR Report Key21975065
Report Number9710602-2025-00810
Device Sequence Number13058438
Product Code BSZ
UDI-Device Identifier00840682146425
UDI-Public(01)00840682146425(11)241030
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K213867
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 05/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA1
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/09/2025
Initial Report FDA Received Date05/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/30/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
SEE H11
Patient Sequence Number1
Patient SexUnknown
-
-