Catalog Number A2114
Medical Device Problem Code Device Slipped (1584)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
World neurosurgery (2020) published an article titled, "intraoperative head slippage with the head clamp system can occur during epileptic surgery." this is 2 of 3 reports for patient #1 of table 2 of the literature and linked to mfg report numbers 3004608878-2025-00084 and 3004608878-2025-00086: background - during neurosurgery, we use a head clamp system for firm head fixation.However, we have encountered slippage using the head clamp system, although this has not been adequately studied.In the present study, to increase the reliability of the analysis using a more homogeneous type of patient data, we conducted a prospective study of patients who had undergone epileptic surgery.We examined the potential risk factors for head slippage and postulated that the location of the pins might be important.Patients - the participants were japanese patients who had been treated from (b)(6) 2019 to (b)(6) 2020 at the comprehensive epilepsy center, seirei hamamatsu general hospital.Malfunction - head slippage had occurred in 3 of the 21 patients (14.3°/o; table 2).The distance between the preoperative and postoperative slipped pins was 2.5 mm.The slipped pins were the upper 2 pins but not the lowest pin in all 3 patients (table 2).However, no complications were associated with head slippage in these 3 patients.Conclusion - we should consider that head clamp slippage could occur intraoperatively.Https://doi.Org/10.1016/j.Wneu.2020.07.051.
 
Additional Manufacturer Narrative
Updated fields: d4, d9, g3, g6, h2, h3, h6, h11.The mayfield infinity xr2 clamp (a2114) was not returned, and no specific device was identified, therefore, failure analysis cannot be completed due to the lack of information received to perform a complete investigation.Additionally, the lot/serial number has not been provided to perform a device history record (dhr) review.As a result, the definite root cause cannot be identified.Based on the reported complaint, the probable root cause is improper or suboptimal placement of the skull clamp on the patient.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD INFINITY XR2 SKULL CLAMP
Common Device Name
MAYFIELD
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH 45227
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 08540
6099362319
MDR Report Key21975840
Report Number3004608878-2025-00085
Device Sequence Number7310352
Product Code HBL
UDI-Device Identifier10381780253761
UDI-Public10381780253761
Combination Product (Y/N)N
PMA/510(K) Number
K130389
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Literature
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberA2114
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/10/2025
Supplement Date Received by Manufacturer06/09/2025
Initial Report FDA Received Date05/07/2025
Supplement Report FDA Received Date06/20/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age19 YR
Patient SexMale
-
-