ST. JUDE MEDICAL AGILIS¿ NXT STEERABLE INTRODUCER DUAL-REACH¿ 61 CM SHEATH, MEDIUM CURL, 8.5F; INTRODUCER, CATHETER Back to Search Results
Model Number G408319
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/24/2025
Type of Reportable Event Malfunction
Event or Problem Description
During the atrial fibrillation procedure, after the puncture the sheath was inserted, but blood leaked from the hemostasis valve.The introducer was replaced and the procedure was completed with no adverse consequences to the patient.
 
Additional Manufacturer Narrative
One 8.5f agilis steerable introducer sheath was received for evaluation.The sheath passed pressure and aspiration leak testing with no anomalies observed.The cap was removed from the hemostasis hub and the hemostasis seals were microscopically inspected.Tearing, resulting in a hole, was noted in the proximal and distal seals.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the damaged seals remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AGILIS¿ NXT STEERABLE INTRODUCER DUAL-REACH¿ 61 CM SHEATH, MEDIUM CURL, 8.5F
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
wendy mccarty
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key21975927
Report Number3008452825-2025-00209
Device Sequence Number7227781
Product Code DYB
UDI-Device Identifier05414734206129
UDI-Public(01)05414734206129(10)10648164(17)271130
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K061363
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG408319
Device Catalogue NumberG408319
Device Lot Number10648164
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2025
Supplement Date Received by Manufacturer06/04/2025
Initial Report FDA Received Date05/07/2025
Supplement Report FDA Received Date06/23/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/13/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-