| Model Number |
PVPK2015L20-A |
| Medical Device Problem Codes |
Stretched (1601); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the spiral guidewire uncoiled when removing from the catheter.No harm to the patient was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Further information about the event and device information has been requested.A supplemental emdr will be sent when the investigation is completed.
|
| |
|
Event or Problem Description
|
|
Manufacturer reference # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation for customer product complaint related to picco catheter which reported guidewire uncoiled during procedure.The faulty part was not returned so far.There was no patient harm due to issue.However, based on the available information and experience it can be concluded that the most probable cause of the issue was need to use excessive force to retract stuck guidewire.There is no indication for a systematic root cause as a deficiency of design, production or material.Overall, investigations did indicate that the device failed to meet its specification when the event occurred.A relationship between the device and the complaint is therefore considered as likely.Upon the complaint occurrence the device was involved and used on a patient for advanced hemodynamic monitoring.Please note, that according to ifu for picco catheter: warning: do not withdraw guide wire against needle bevel to avoid possible severing or damage of guide wire.Warning: if resistance is encountered when removing the guide wire, the guide wire could be kinked about tip of catheter within vessel (fig.K).In this case withdraw the catheter by 2 to 3 cm and try again to remove the guide wire.Use of excessive force may tear off the guide wire.Therefore, guide wire and catheter are to be removed simultaneously.
|
| |
|
Search Alerts/Recalls
|