Catalog Number 121932150
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/16/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the pinnacle liner was damaged in situ, and the patient needed revision as a result.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #
=
>(b)(4).Investigation summary
=
> no device associated with this report was received for examination.A search of the j&j medtech orthopaedics nonconformance (nc) quality system found no nc¿s associated with this product 121932150/lot m28m33combination.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot
=
> a search of the j&j medtech orthopaedics nonconformance (nc) quality system found no nc¿s associated with this product 121932150/lot m28m33combination.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d9, d10 (concomitant) if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: possible product failure, pinnacle liner damaged in situ and needing revision.Surgeon would like an analysis of the product to determine mode of failure.The product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device did found that the pinn mar +4 10d 32idx50od has three of six anti-rotation device (ard) tabs sheared off and they were not returned for evaluation.Additionally, the liner appeared to have been edge loaded causing eccentric deformation in the rim of the device and ultimately contributing to the failure of the anti-rotation devices (ards).There are machining lines from the manufacturing process yet visible within the liner which would not be as obviously present if would be more centrally loaded.Due to the observed condition of the involved implants it is reasonable to conclude that a disassociation event occurred between the liner and the cup.There is no indication that a design or manufacturing issue has caused or contributed to the reported event.Although the liner failure cannot be traced to design or manufacturing, a definite root cause cannot be established due to there being multiple factors that may influence.Consideration must be given to all potential influences such as surgical process, patient variables (e.G.Activity level and use), anatomical considerations and patient changes over time.Any conclusions from the investigational input provided have to be placed into context with all other relevant factors.It is highly recommended to the patient to consult with their healthcare professional for further assessment.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the pinn mar +4 10d 32idx50od would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation was performed for the finished device m28m33 number, and no non-conformances were identified during manufacturing.H11 additional narrative: corrected: h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MAR +4 10D 32IDX50OD
Common Device Name
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key21976222
Report Number1818910-2025-07245
Device Sequence Number9197375
Product Code LPH
UDI-Device Identifier10603295014041
UDI-Public(01)10603295014041(17)280229(10)M28M33
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K033273
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number121932150
Device Lot NumberM28M33
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/16/2024
Supplement Date Received by Manufacturer05/15/2025
08/28/2025
11/11/2025
Initial Report FDA Received Date05/07/2025
Supplement Report FDA Received Date05/16/2025
09/02/2025
11/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/15/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DELTA CER HEAD 12/14 32MM +5.; PINNACLE SECTOR II CUP 50MM.; UNK HIP FEMORAL HEAD CERAMIC.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR