| Catalog Number |
0102L16 |
| Medical Device Problem Code |
Component Misassembled (4004)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/21/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the nurse and nursing student attempted to insert foley catheter into patient due to urine retention.After not receiving urine return with insertion the nurse called in writer to assist with insertion.Writer tried repositioning the catheter with no returns.Foley catheter removed to attempt with a new catheter.Once removed writer and original nurse identified that there were no drainage holes in the end of the catheter.New catheter inserted with no complications.Unit nurse educator notified of situation.Foley catheter with no holes kept on unit to help with possible manufacturer issues.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because no sample was returned was returned for evaluation.It was unknown whether the device had met relevant specifications.A potential root cause for this failure mode could be eye slug not removed.The dhr review could not be performed without a lot number.The instructions for use were found adequate and states the following: warning: do not use ointments or lubricants having a petrolatum base.They will damage latex.Visually inspect the product for any imperfections or surface deterioration prior to use.Use luer tip syringe to inflate with stated ml of sterile water.For pre-filled products, remove clip and squeeze reservoir to inflate with stated ml of sterile water.For urological use only.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.This is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the nurse and nursing student attempted to insert foley catheter into patient due to urine retention.After not receiving urine return with insertion the nurse called in writer to assist with insertion.Writer tried repositioning the catheter with no returns.Foley catheter removed to attempt with a new catheter.Once removed writer and original nurse identified that there were no drainage holes in the end of the catheter.New catheter inserted with no complications.Unit nurse educator notified of situation.Foley catheter with no holes kept on unit to help with possible manufacturer issues.
|
| |
|