• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STERIS MEXICO, S. DE R.L. DE C.V. AMSCO; AMSCO 600 STEAM STERILIZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS MEXICO, S. DE R.L. DE C.V. AMSCO; AMSCO 600 STEAM STERILIZER Back to Search Results
Model Number AMSCO 600 STEAM STERILIZER
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/08/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation of this event is in progress.A follow-up report will be submitted when additional information becomes available.
 
Event or Problem Description
The user facility reported that water was leaking from their amsco 600 sterilizer onto the floor.No report of injury.
 
Additional Manufacturer Narrative
A steris service technician arrived onsite following the reported event to inspect the amsco 600 sterilizer.The technician found that the safety valve was not operating properly subsequently causing the reported event.The safety valve was returned to the supplier for evaluation and found that one of the internal components of the safety valve was misaligned.The cause of the misalignment could not be determined.The technician replaced the safety valve, tested the function and operation of the sterilizer and returned the device to service.A 3-year complaint review confirmed this to be an isolated event.No additional issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMSCO
Common Device Name
AMSCO 600 STEAM STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
MDR Report Key21978318
Report Number3005899764-2025-00037
Device Sequence Number7104850
Product Code FLE
UDI-Device Identifier00724995179427
UDI-Public(01)00724995179427(11)250117
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAMSCO 600 STEAM STERILIZER
Device Catalogue NumberEF1201242
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/08/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/07/2025
Supplement Report FDA Received Date09/08/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/17/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-