• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON® FENESTRATED SCREW SET; BONE CEMENT, POSTERIOR SCREW AUGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON® FENESTRATED SCREW SET; BONE CEMENT, POSTERIOR SCREW AUGMENTATION Back to Search Results
Model Number 55750026545
Medical Device Problem Code Backflow (1064)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having percutaneous pe dicle screw (pps) fixation therapy for l1 compression fracture.It was reported that the cement was refilled with fns, but it was not delivered to the vertebral body and leaked into the screw head.Both places were right, th12 and l2.It was not possible to obtain the fixation properties that were originally expected.Cement remained in the screw head, making it impossible to install the rod.Th12 was removed and the screw was re-placed.L2 seemed to have been able to remove the cement at first glance, but because the set screw could not be tightened completely due to the remaining cement near the crown, the set screw burst open during the final tightening, so the v5 tab was completely removed and the rod was re-placed.The set screw did not fit into the screw head and came loose during the final fastening.There was a delay of less than 60 minutes in overall procedure.There were no patient symptoms reported.There were no further complications reported regarding the event.
 
Additional Manufacturer Narrative
H3: product analysis # (b)(4), part # 55750026545, lot # h5749806, visual inspection confirmed the cement has leaked in the head of the screw.The cement leak appears to be from the failed attempt in delivering the cement in the vertebral body.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON® FENESTRATED SCREW SET
Common Device Name
BONE CEMENT, POSTERIOR SCREW AUGMENTATION
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer (Section G)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key21979117
Report Number1030489-2025-01916
Device Sequence Number12988799
Product Code PML
UDI-Device Identifier00763000148836
UDI-Public00763000148836
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K201362
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number55750026545
Device Catalogue Number55750026545
Device Lot NumberH5749806
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/14/2025
Supplement Date Received by Manufacturer06/25/2025
Initial Report FDA Received Date05/07/2025
Supplement Report FDA Received Date07/16/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1