• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET SPINE, INC. PLUG; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE, INC. PLUG; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Catalog Number 2000-1005
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 04/06/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery was performed to remove an ha coated pedicle screw construct from levels l1 through s1 because the hardware was causing the patient pain.The patient had solidly fused prior to the revision so the construct was no longer of use and was not replaced with any other hardware.One of the screws on the bottom of the construct was unable to be removed because the tip of the height adjuster broke off in the screw head.This is report eight of twenty six for this event.
 
Additional Manufacturer Narrative
D4: the remainder of the udi number is unknown because the lot number is not available.Without a product return, no product evaluation is able to be conducted.The lot number is unknown; therefore the device history records are unable to be reviewed.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.This report references the same patient as report 3012447612-2025-00177.Reference reports 3012447612-2025-00178 through 3012447612-2025-00203 for this event.
 
Event or Problem Description
It was reported that a revision surgery was performed to remove an ha coated pedicle screw construct from levels l1 through s1 because the hardware was causing the patient pain.The patient had solidly fused prior to the revision so the construct was no longer of use and was not replaced with any other hardware.One of the screws on the bottom of the construct was unable to be removed because the tip of the height adjuster broke off in the screw head.This is report eight of twenty six for this event.
 
Additional Manufacturer Narrative
Additional information in h6: investigation type, findings, and conclusions.Device evaluation: the device was not returned and no photos were provided, so an evaluation is unable to be performed.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to unknown patient or surgical factors.The ifu states that the construct should be removed after healing is complete.Constructs that are not removed can cause patient pain.Dhr review: the lot number was not reported and the device was not returned, therefore a dhr review is unable to be performed.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLUG
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer Contact
teresa george
10225 westmoor dr.
westminster, CO 80021
3035018506
MDR Report Key21981100
Report Number3012447612-2025-00185
Device Sequence Number8907988
Product Code MNH
UDI-Device Identifier00880304925816
UDI-Public(01)00880304925816
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K061441
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number2000-1005
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/07/2025
Supplement Report FDA Received Date07/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention; Hospitalization;
Patient SexFemale
Patient EthnicityNon Hispanic
-
-