Catalog Number CXT201412
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Vascular Dissection (3160)
Date of Event 04/17/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.H6: code d12: according to the gore® excluder® conformable aaa endoprosthesis instructions for use (ifu), adverse events that may occur and/or require intervention include, but are not limited to: dissection, perforation, or ruptures of the aortic vessel and surrounding vasculature w.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may have not been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
The following information was reported to gore: on (b)(6) 2025, the patient underwent an endovascular treatment for an abdominal aortic aneurysm using gore® excluder® conformable aaa endoprostheses and gore® excluder®aaa endoprostheses.A gore® molding & occlusion balloon catheter (mob) was used as an accessory.The trunk-ipsilateral leg endoprosthesis was placed below the left renal artery.A gore® molding & occlusion balloon catheter (mob) was used for balloon molding.Final angiography revealed a retrograde dissection from just above the renal artery/proximal end of the device to the diaphragm area.The patient was decided to be monitored, and the procedure was completed.The physician mentioned that the patient proximal neck was narrow, and the balloon molding may have been too forceful.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 2 B/P
32470 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
rie kaba
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key21981671
Report Number3007284313-2025-03957
Device Sequence Number12904699
Product Code MIH
UDI-Device Identifier00733132650996
UDI-Public(01)00733132650996(17)261202
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P200030
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberCXT201412
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/17/2025
Initial Report FDA Received Date05/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/04/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age77 YR
Patient SexFemale
-
-