H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.H6: code d12: according to the gore® excluder® conformable aaa endoprosthesis instructions for use (ifu), adverse events that may occur and/or require intervention include, but are not limited to: dissection, perforation, or ruptures of the aortic vessel and surrounding vasculature w.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may have not been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
The following information was reported to gore: on (b)(6) 2025, the patient underwent an endovascular treatment for an abdominal aortic aneurysm using gore® excluder® conformable aaa endoprostheses and gore® excluder®aaa endoprostheses.A gore® molding & occlusion balloon catheter (mob) was used as an accessory.The trunk-ipsilateral leg endoprosthesis was placed below the left renal artery.A gore® molding & occlusion balloon catheter (mob) was used for balloon molding.Final angiography revealed a retrograde dissection from just above the renal artery/proximal end of the device to the diaphragm area.The patient was decided to be monitored, and the procedure was completed.The physician mentioned that the patient proximal neck was narrow, and the balloon molding may have been too forceful.
|