INTEGRUM AB OPRA IMPLANT SYSTEM; AXOR II Back to Search Results
Model Number 1288
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Fall (1848)
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2025: integrum received a report form stating that the axor ii i9815 had fallen off from the abutment.The patient fell which caused an internal bleeding in the stump and hospitalization.The exact date of the incident is unknown.Manufacturing investigation performed, the unit has been manufactured according to specification.The unit not yet received; hence no technical investigation has been performed yet.
 
Event or Problem Description
2025-04-24: integrum received a report form stating that the axor ii (b)(6) had fallen off from the abutment.The patient fell which caused an internal bleeding in the stump and hospitalization.The exact date of the incident is unknown.Manufacturing investigation performed, the unit has been manufactured according to specification.The unit not yet received, hence no technical investigation has been performed yet.2025-06-13: technical investigation of unit (b)(6) performed.During the investigation the device functions were tested and components inspected.The reported issue could not be replicated.The device is within the manufacturers specifications and functions as designed.The unit showed signs of poor cleaning and the clamps had marks indicating that the abutment may not have been fully inserted at some point.Integrum has informed the clinic on the importance of donning the axor ii according to the ifu to ensure a stable connection and prevent damage or wear to the clamps.As stated in the ifu, the device shall not be used if a stable connection cannot be achieved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPRA IMPLANT SYSTEM
Common Device Name
AXOR II
Manufacturer (Section D)
INTEGRUM AB
gemenskapens gata 9
molndal, 43153
SW  43153
Manufacturer (Section G)
INTEGRUM
gemenskapens gata 9
molndal, 43153
SW   43153
Manufacturer Contact
susanne börjesson
gemenskapens gata 9
molndal, 43153
SW   43153
MDR Report Key21983920
Report Number3011386779-2025-00231
Device Sequence Number4263770
Product Code PJY
UDI-Device Identifier07340152100634
UDI-Public07340152100634
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
P190009
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Remedial Action Inspection
Type of Report Initial,Followup
Report Date (Section B) 06/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number1288
Device Lot Number61893
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2025
Supplement Date Received by Manufacturer06/12/2025
Initial Report FDA Received Date05/08/2025
Supplement Report FDA Received Date06/16/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient SexUnknown
-
-