AIZU OLYMPUS CO., LTD. TRACHEAL INTUBATION FIBERSCOPE Back to Search Results
Model Number LF-TP
Medical Device Problem Code Microbial Contamination of Device (2303)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/31/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the scope tested positive for an unexpected contamination.The issue was found during a routine culture of the scope.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
 
Additional Manufacturer Narrative
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the completed investigation's review of the customer cleaning disinfection and sterilization (cds) processes, results of third-party testing, the device evaluation and the final investigation.The complaint investigation reviewed the customer provided the cds processes where [no obvious deviations from instructions for use (ifu) were identified.Olympus provided the following result of the culture test, performed at the third-party labs: sampling date: (b)(6) 2025.Sampling from: sampling solution instrument / biopsy / suction channel; swab distal end cfu: 0 cfu.Bacterial identification: n/a.Sampling date: (b)(6) 2025.Sampling from: sampling solution instrument / biopsy / suction channel; swab distal end cfu: 0 cfu.Bacterial identification: n/a.The device was evaluated where no abnormalities were found that could have led to the reported event.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the ifu before repair, the results conformed to the regulation's recommendation.
 
Event or Problem Description
No additional information received from customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRACHEAL INTUBATION FIBERSCOPE
Common Device Name
TRACHEAL INTUBATION FIBERSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key21984590
Report Number9610595-2025-08072
Device Sequence Number12899913
Product Code EOQ
UDI-Device Identifier04953170136825
UDI-Public04953170136825
Combination Product (Y/N)N
PMA/510(K) Number
K981543
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 08/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLF-TP
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/10/2025
Supplement Date Received by Manufacturer07/14/2025
Initial Report FDA Received Date05/08/2025
Supplement Report FDA Received Date08/05/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/15/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-