| Model Number |
LF-TP |
| Medical Device Problem Code |
Microbial Contamination of Device (2303)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/31/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the scope tested positive for an unexpected contamination.The issue was found during a routine culture of the scope.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the completed investigation's review of the customer cleaning disinfection and sterilization (cds) processes, results of third-party testing, the device evaluation and the final investigation.The complaint investigation reviewed the customer provided the cds processes where [no obvious deviations from instructions for use (ifu) were identified.Olympus provided the following result of the culture test, performed at the third-party labs: sampling date: (b)(6) 2025.Sampling from: sampling solution instrument / biopsy / suction channel; swab distal end cfu: 0 cfu.Bacterial identification: n/a.Sampling date: (b)(6) 2025.Sampling from: sampling solution instrument / biopsy / suction channel; swab distal end cfu: 0 cfu.Bacterial identification: n/a.The device was evaluated where no abnormalities were found that could have led to the reported event.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the ifu before repair, the results conformed to the regulation's recommendation.
|
| |
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|