| Catalog Number |
136504000 |
| Medical Device Problem Code |
Corroded (1131)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
(b)(6) first report of retrievals for statement of work 10 for q1 of 2025.(b)(6) is the research location, and not the facility of placement.A depuy synthes implant was revised and reviewed for analysis reason for revision: removal of prostalac spacer after infection treatment.Reimplantation of final implants.At removal the femoral head showed corrosion.At removal the acetabular liner showed minor yellowing.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.D4: udi: as the catalog/model number was not provided, the (01)gtin is not available.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not performed.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: quantity manufactured: (b)(4).Date of manufacture: 10/21/2022.Any anomalies or deviations identified in dhr: none.Expiry date: 9/30/2027.Ifu-0902-00-701.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances / manufacturing irregularities were identified during manufacturing.
|
| |
|
Search Alerts/Recalls
|