| Model Number |
CCA0T0 |
| Medical Device Problem Code |
Failure to Unfold or Unwrap (1669)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that during an intraocular lens (iol) implant procedure, trailing haptic was stuck to posterior side of optic noticed upon insertion in the eye.Haptic was released from the optic.Lens remains implanted.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in h.3., h.6.And h.11.The product was not returned for analysis.It was reported that haptic was observed to be stuck to the optic after implantation.The root cause of the reported issue is deemed to be related to manufacturing process change that increased the likelihood of company haptics adhering to the posterior optic surface following delivery with the company device.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
|
| |
|