| Catalog Number |
TSVMWB11 |
| Medical Device Problem Codes |
Fracture (1260); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Inflammation (1932); Pain (1994)
|
| Date of Event |
12/05/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).A4: patient weight: unknown / not provided, d4: additional device information: unknown / not provided, g4: premarket identification pma/510(k) #: k101880.
|
| |
|
Event or Problem Description
|
|
Doctor reported that implant at tooth site 37 was removed due to infection and implant fracture.Pain and inflammation were reported as a result of the event.Doctor performed cleaning and augmentation for a later implantation.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).
|
| |
|
Event or Problem Description
|
|
Additional information received.Doctor confirmed the implant was not fractured, a fractured abutment hex was found stuck inside the implant.Doctor reported that implant at tooth site 37 was removed due to infection.Pain and inflammation were reported as a result of the event.Doctor performed cleaning and augmentation for a later implantation.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report.B5: describe event or problem.D4: additional device information.D9: device availability.G3: date received by manufacturer.G6: type of report.H1: type of reportable event.H2: follow up type.H3: device evaluated by manufacturer.H4: device manufacturer date.H6: adverse event problem.H10: additional narrative.Dhr review was completed for the subject lot number 63727036.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Sterilization record was reviewed and verified to have passed all sterilization activities with no issues or nonconformities identified.Complaint history review was performed for the reported lot number 63727036 for similar events and no other complaint was identified.Review completed utilizing keywords: dental: medical: infection.A summary investigation has been completed for infection events identifying that a definitive root cause cannot be identified due to the wide range of external (non-design/ non-manufacturing related) factors potentially impacting the sterility of the implant and its environment.Should additional information be received which indicates that the device may have caused or contributed to the infection event, an additional report will be submitted.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|