| Catalog Number |
8065921540 |
| Medical Device Problem Code |
Dull, Blunt (2407)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
05/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported that before surgery the ophthalmic knife was damaged and bent.The procedure and patient details were not reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in b.5 the manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
Additional information confirmed that the scheduled procedure was phacoemulsification surgery.The surgery was completed by using other side of the port by an alternate system.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information is provided in sections d.9, h.3, h.6 and h.11.One opened knife was received for the report of damage (bent).Sample was visually inspected and found to be nonconforming with a bent blade at the connection of the blade and the handle.Penetration testing could not be performed due to the damage of the sample.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The investigation conducted a non-conformance review of the reported lot number.No deviations were identified and all corresponding production release specifications defined in the device master record were met.Sample was visually inspected and found to be nonconforming with a bent blade at the connection of the blade and the handle.The damage to the returned sample is consistent with damage that can occur when the blade contacts a hard surface such as the protective blade tray when product is improperly removed or inserted after use, or improper handling, or contact with another instrument during surgery or set-up.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.All knives are 100% inspected by trained operators using a minimum of 10x magnification during manufacturing.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all alcon products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
|
| |
|