This is 2 of 3 reports (different patients, same product family) linked to mfg report numbers: 3013886523-2025-00092.3013886523-2025-00096.A physician reported: on an unspecified date, a patient was implanted with a codman hakim programmable valve.On an unknown date, the shunt failed, requiring the doctor to perform additional surgery to replace it.Outcome of the event was unknown.Causality of the event was reported as possible with codman hakim programmable valve.The patient's medical history was not reported.No concurrent condition was reported.No family history was reported.Past medication was not reported.Concomitant medication and co-suspect medications were not reported.No drug allergies were reported.Pertinent lab data was not reported.
|
|
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h11.The hakim valve (id 823148) was not returned for evaluation after three good faith attempts (gfes) were made.Lot number information has been provided; therefore, device history record (dhr) was reviewed, and no anomalies were found.The root cause of the reported issue could not be determined; however, the possible root cause for the issue reported by the customer could be due to biological debris and protein build up interfering with the valve.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|