| Model Number |
1000096 |
| Medical Device Problem Code |
Pumping Stopped (1503)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Per the tandem user guide: "do not use your pump if you think it might be damaged due to dropping it or hitting it against a hard surface." no product was returned for evaluation.Should new relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a malfunction alarm occurred after the pump was dropped.During troubleshooting with technical support, the malfunction alarm was cleared, and insulin delivery was resumed successfully.There was no adverse impact to the customer¿s blood glucose level.
|
| |
|