Catalog Number 5743690
Medical Device Problem Codes Fracture (1260); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/19/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
On (b)(6) 2025, a patient underwent hemosplit catheter insertion for unknown medical condition.During insertion the catheter broke outside the blood vessel, and the broken part did not enter the vascular system.The procedure was completed by replaced with another device.There was no reported patient injury.
 
Event or Problem Description
On (b)(6) 2025, a patient underwent hemosplit catheter insertion for unknown medical condition.During insertion the catheter broke outside the blood vessel, and the broken part did not enter the vascular system.The procedure was completed by replaced with another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H11: manufacturing review: the device history records have been reviewed and this lot met all release criteria.Investigation summary: although the physical device was not returned for evaluation, one photo was provided for review.The photo shows a hemosplit catheter in which the catheter is split into two segments one segment from proximal end up to the catheter cuff and second segment consists of middle section post cuff up to the distal catheter tip.Therefore, the investigation is confirmed for the reported catheter fracture and identified material separation issues.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.D4 (unique identifier (udi) #), g3, h4, h6 (device, method, result).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOSPLIT CATHETER
Common Device Name
DIALYSIS CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key21991982
Report Number3006260740-2025-02965
Device Sequence Number19905404
Product Code MSD
UDI-Device Identifier00801741013263
UDI-Public(01)00801741013263(11)241127(17)260331(10)REJS0829
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K030020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2026
Device Catalogue Number5743690
Device Lot NumberREJS0829
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/15/2025
Supplement Date Received by Manufacturer06/24/2025
Initial Report FDA Received Date05/09/2025
Supplement Report FDA Received Date07/17/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/27/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1