| Catalog Number |
5743690 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025, a patient underwent hemosplit catheter insertion for unknown medical condition.During insertion the catheter broke outside the blood vessel, and the broken part did not enter the vascular system.The procedure was completed by replaced with another device.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025, a patient underwent hemosplit catheter insertion for unknown medical condition.During insertion the catheter broke outside the blood vessel, and the broken part did not enter the vascular system.The procedure was completed by replaced with another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: the device history records have been reviewed and this lot met all release criteria.Investigation summary: although the physical device was not returned for evaluation, one photo was provided for review.The photo shows a hemosplit catheter in which the catheter is split into two segments one segment from proximal end up to the catheter cuff and second segment consists of middle section post cuff up to the distal catheter tip.Therefore, the investigation is confirmed for the reported catheter fracture and identified material separation issues.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.D4 (unique identifier (udi) #), g3, h4, h6 (device, method, result).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|