| Model Number |
MMT-975A |
| Medical Device Problem Code |
Device Handling Problem (3265)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1: patient city: (b)(6).Patient country: united arab emirates.
|
| |
|
Event or Problem Description
|
|
Reference number (b)(4).Event occurred in the united arab emirates.It was reported that the patient experienced that infusion set cannula was fired before insertion which occurred on (b)(6) 2025.No further information was available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information - this mdr is being submitted to include the below: h6: investigation results under type of investigation, investigation findings, investigation conclusions.H11: investigation summary: revision 21 of (b)(4) does not require a complaint that is type 2 reportable to open a child investigation.This child investigation was opened against a previous revision of (b)(4).Complaint investigation results: a complaint investigation has been initiated under complaint investigation child record (b)(4).The batch 6003343, in question was manufactured at the reynosa site.Device history record (dhr) review: the lot 6003343 was manufactured according to the work instruction (wi) version 69 and manufactured in the line li103 on 29-jan-2024, with a total of (b)(4) units.Review of the dhr showed that all relevant tests required during the related processes had been fulfilled and met the requirements.No deviation was identified, nor maintenance events were recorded related to complaint code.Conclusion summary of complaint investigation: as a result of the following: no non-conformance (nc) raised during production related to complaint code, no trend identified for the lot in question and malfunction code, no further actions are required.This complaint will not require further root cause investigation nor corrective and preventive action (capa) plan.Therefore, this issue will be monitored through the post market surveillance activities.
|
| |
|
Event or Problem Description
|
|
To date no additional patient or event details have been received.
|
| |
|