MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH SONATA; COCHLEAR IMPLANT Back to Search Results
Model Number SONATA
Medical Device Problem Code Impedance Problem (2950)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/07/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event or Problem Description
Although the family did not report any specific trauma, they said that the child is very active and could have been hit his head without the family seeing.During the session, the child never stopped and tended to climb things all the time.Both side affected channels were detected on the same day.Re-implantation is considered.
 
Additional Manufacturer Narrative
Additional information: based on the received information, a damage to the active electrode due to excessive mechanical stress seems very likely.However to determine an exact root cause device investigation would be necessary.Re-implantation is considered but no date has been scheduled yet.
 
Event or Problem Description
Although the family did not report any specific trauma, they said that the child is very active and could have been hit his head without the family seeing.During the session, the child never stopped moving and tended to climb on things all the time.Re-implantation is considered.
 
Event or Problem Description
Although the family did not report any specific trauma, they said that the child is very active and could have been hit his head without the family seeing.During the session, the child never stopped moving and tended to climb on things all the time.The user has been re-implanted.
 
Additional Manufacturer Narrative
Additional information: based on the received information, a damage to the active electrode due to excessive mechanical stress seems very likely.However to determine an exact root cause device investigation would be necessary.The concerned device was explanted but has not been received for investigation yet.
 
Event or Problem Description
Although the family did not report any specific trauma, they said that the child is very active and could have been hit his head without the family seeing.During the session, the child never stopped moving and tended to climb on things all the time.The recipient has been re-implanted.
 
Additional Manufacturer Narrative
Conclusion: overload fractures in the active electrode which are consistent with an external mechanical impact were determined to be the root cause of device failure.The active electrode also showed some additional damage to the electrode wires, as being caused by minute device mobility.An accident or impact might have weakened the fixation of the implant, consequently leading to electrode mobility.The problems given in the recipient report appear to match the damage found.Other damages found during device investigation are most likely related to the explantation surgery.This is a final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SONATA
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key21992593
Report Number9710014-2025-00416
Device Sequence Number20911079
Product Code MCM
UDI-Device Identifier09008737049133
UDI-Public(01)09008737049133
Combination Product (Y/N)N
Initial Reporter CountryTU
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 08/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/04/2024
Device Model NumberSONATA
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/07/2025
Supplement Date Received by Manufacturer05/07/2025
05/07/2025
05/07/2025
Initial Report FDA Received Date05/09/2025
Supplement Report FDA Received Date05/22/2025
06/12/2025
08/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age5 YR
Patient SexMale
-
-