It was reported a patient underwent an unknown procedure on an unknown date and a reservoir was used.When trying to connect the connector and the bag before use with the patient, it felt like the connector was broken, so the use was stopped.Another product was used to complete the case.There were no adverse consequences to the patient.Further details are not provided.Additional information has been requested.
|
|
Product complaint:(b)(4).Additional information has been requested and received.Attempts to obtain the device have been made.If further details are received at a later date a supplemental medwatch will be sent.? what is the lot number? => j2400398 ? please perform and document the follow up attempt for product return.Please document the shipment tracking number in notes or rmao sections.=>the device has been received at sukagawa and will be shipped.Please check rmao.? please provide the source or name and title of external person providing answers to follow-up (external person submitting answers to sales rep or loc/bq) =>hiromi tokuyama h3 evaluation: product sample received for analysis.Complaint sample review : one complaint sample of reservoir bulb with partial connection port was received for evaluation, product was inspected visually, and found that there were significant cut marks visible on the cut-off surface of the bulb port.Documents review : during investigation of complaint, batch review of in-process and final packaging inspections, as well as the manufacturing process is performed, these products are reported to be manufactured, inspected, and packaged in conformance with customer's specifications and have been found to be acceptable within all parameters.No discrepancy as per the reported defect is observed.Retention sample review : no negative observation was found.Review of defective sample's image / video : not applicable, as there was no complaint sample image / video received for evaluation.Based on above observations, it is suspected that the product might have handled differently at user end, and lead to the defect generation.As per standard practice, 100% functional test and 100% visual inspection was carried out, before and after packing of finished goods, prior to the product release.There was no scope to miss such defect, at manufacturing / release stage.The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|