• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TPRLC XR T1 PPS 15X150MM; PROSTHESIS, HIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TPRLC XR T1 PPS 15X150MM; PROSTHESIS, HIPS Back to Search Results
Model Number N/A
Medical Device Problem Code Packaging Problem (3007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/16/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).D10: 51-104150 tprlc 133 t1 pps ho 15x150mm (b)(4).51-105160 tprlc xr t1 pps 16x152mm (b)(6).51-106160 tprlc 133 mp type1 pps so 16.0 (b)(6).51-145160 tprlc xr mp t1 pps 16x117mm (b)(6) 51-103170 tprlc 133 t1 pps so 17x154mm (b)(6) 51-103130 tprlc 133 t1 pps so 13x146mmtp (b)(6) 51-104140 tprlc 133 t1 pps ho 14x148mm (b)(6).G2: foreign: japan.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported during inspection that the sterile packaging of the circulated items was damaged.There was no patient involvement.
 
Event or Problem Description
No additional event information to report at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.Updated: g3, g6, h2, h3, h6, h10.Visual evaluation of the returned product/provided photos identified damage to the sterile packaging (blister).Sterility has not been compromised.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The reported event did not occur in an operating room or as part of a medical procedure; medical records are not available for review.The condition of the device when it left zimmer biomet is considered conforming to specification.The root cause of the reported event can be attributed to transit damage and a packaging design issue.The reported event has been confirmed by evaluation of the returned product and provided photos.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC XR T1 PPS 15X150MM
Common Device Name
PROSTHESIS, HIPS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key21994566
Report Number0001825034-2025-01357
Device Sequence Number11828017
Product Code LPH
UDI-Device Identifier00880304517165
UDI-Public(01)00880304517165(17)301023(10)6881901
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K120030
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number51-105150
Device Lot Number6881901
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2025
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/16/2025
Supplement Date Received by Manufacturer08/05/2025
Initial Report FDA Received Date05/09/2025
Supplement Report FDA Received Date08/13/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/23/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE.
Patient SexUnknown
-
-