MAUDE Adverse Event Report: ETHICON INC. VCL+ VIO 27IN 3-0 S/A SH-1; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VCL+ VIO 27IN 3-0 S/A SH-1; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number VCP311H
Medical Device Problem Codes Break (1069); Product Quality Problem (1506); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an unknown orthopedic procedure in (b)(6) 2025 and suture was used.It was reported that the orthopaedic surgeon mentioned that there are 3 sutures that broke when he was doing the suturing.The needle detached from the swage, and the suture that he had tied opened on its own and before not secure.There were no patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 5/8/2025.H6 component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, the following was obtained: - it was reported that three vicryl plus sutures broke during suturing, with the needle detaching from the swage, and one suture tying itself open before being secured.Can you clarify if these three issues happened with each of the three devices, or did each problem occur with each suture? ans: the three sutures happened on the same patient.Same procedure.There were three sutures affected.According to what the surgeon mentioned.- please provide the source or name of person providing answers to follow-up questions: ans: sales rep was not present in the case.It was being told by the surgeon himself and he was asking if j&j changed raw material, changed factory etc as this never happened before in his practice.And he is a senior surgeon.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL+ VIO 27IN 3-0 S/A SH-1
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-GMBH
robert-koch strasse 1
p.o. box 1409
norderstedt
GM  
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key21994993
Report Number2210968-2025-05156
Device Sequence Number8312830
Product Code GAM
UDI-Device Identifier10705031052192
UDI-Public(01)10705031052192(17)270630(10)UHBDHPW0
Combination Product (Y/N)Y
Initial Reporter CountryMY
PMA/510(K) Number
K132580
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberVCP311H
Device Lot NumberUHBDHPW0
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/12/2025
Initial Report FDA Received Date05/09/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/25/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-