VANTIVE US HEALTHCARE LLC LOCKING TITANIUM ADAPTER FOR PERITONEAL DIALYSIS CATHETER; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE Back to Search Results
Catalog Number 5C4129
Medical Device Problem Code Disconnection (1171)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B3/d10: the event occurred on an unspecified date in april 2025, reported as "last week." e1: initial reporter facility name: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that there was a connection issue between the titanium adapter and non-vantive catheter; further described as "the titanium connector detached from the catheter." this occurred four days after a catheter repair procedure in which a new titanium adapter was installed.The repair protocol was repeated and a new titanium adapter was installed.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
Additional information: h6 and h11.H11: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was discarded; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
H11: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Leak testing, functional testing, and assembly of the titanium adapter and sleeve was performed, and no issues were observed.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING TITANIUM ADAPTER FOR PERITONEAL DIALYSIS CATHETER
Common Device Name
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
VANTIVE US HEALTHCARE LLC
round lake IL
Manufacturer (Section G)
VANTIVE HEALTHCARE - SWINFORD
foxford road
swinford, county mayo
EI  
Manufacturer Contact
25212 w. il route 120
round lake, IL 60073
2242706717
MDR Report Key21995984
Report Number9612168-2025-00009
Device Sequence Number4312026
Product Code KDJ
UDI-Device Identifier00085412007403
UDI-Public00085412007403
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K152675
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number5C4129
Device Lot Number14G07H35
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/14/2025
Supplement Date Received by Manufacturer06/04/2025
10/29/2025
Initial Report FDA Received Date05/09/2025
Supplement Report FDA Received Date06/10/2025
11/06/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NON-VANTIVE CATHETER.
Patient SexUnknown
-
-