Continued d2a: biliary catheter for stone removal that may also allow for irrigation and contrast injection.Investigation evaluation: the product said to be involved was returned in an open pouch from the lot number provided in the report.The label matches the product returned.Our laboratory evaluation of the product said to be involved confirmed the report.The device was returned with the syringe still attached to the inflation port.The balloon was ruptured.Under magnification, it was identified there was a portion where the balloon material did not match up and a portion appears to be missing.The missing portion of the balloon was not returned with the device.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: our evaluation of the returned device confirmed the report.A definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.In the report it does state it is unknown if the balloon was tested prior to advancement down the endoscope accessory channel.The ifu states: "verify balloon integrity prior to use by attaching the enclosed syringe to the stopcock and inflating the balloon with air only." a pinhole, split, or rupture in the balloon can occur if the balloon material has come into contact with a sharp object, such as a sharp stone or possibly a burr in the endoscope channel.A split or rupture in the balloon material can also occur if added pressure was applied during extraction.The instructions for use direct the user to "gently withdraw the inflated balloon toward the papilla." the instructions for use contain the following: ¿warning: do not exert excessive pressure on ampulla while extracting stones.If stone does not pass easily, reassess need for sphincterotomy.¿ prior to distribution, all fusion quattro extraction balloon are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: based on the quality engineering risk assessment no corrective action is warranted at this time.A review of the complaint history was conducted.Based on this review, the likelihood of this type of report is rare.Quality assurance will continue to monitor for complaint trends.
|
|
During an endoscopic biliary stone removal, the physician used a cook fusion quattro extraction balloon.It was reported [that] the user attempted to inflate [balloon] with air, but the balloon did not inflate.The use of the device was discontinued, and a new device (unknown lot) was used and the procedure was performed.There was no reportable information at this time.The device was received for evaluation on 15apr2025.It was determined that there is a missing section of the balloon material [subject of this report].A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|