G4: pma/510(k) #: preamendment.H3: device has not yet been returned to manufacturer; however, our investigation is ongoing, and device evaluation summary will be included in our follow up report once our investigation has been completed.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
During a stone removal procedure, the stents from two filiform double pigtail ureteral stent sets were unable to advance over the wireguide.The procedure was completed by using a universa soft stent.The two stents that were unable to advance over the wireguide had different lot numbers.Reference patient identifier: (b)(6) (this report) for stent # 2, and reference patient identifier: (b)(6) for the stent # 1.A third stent was attempted to be used; however, the surgeon mistakenly used a stent that was too short, then removed it.The procedure was complete by using a 4th stent which was a universa soft stent.As section of the device did not remain inside the patient¿s body.The patient did not require any additional intervention due to this occurrence.No patient adverse effects have been reported.
|
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Investigation ¿ evaluation.During a stone removal procedure, the stents from two filiform double pigtail ureteral stent sets were unable to advance over the wireguide.The procedure was completed by using a universa soft stent.The two stents that were unable to advance over the wireguide had different lot numbers.Reference patient identifier (b)(6) (this report) for stent # 2, and reference patient identifier (b)(6) for the stent # 1.A third stent was attempted to be used; however, the surgeon mistakenly used a stent that was too short, then removed it.The procedure was complete by using a 4th stent which was a universa soft stent.As section of the device did not remain inside the patient¿s body.The patient did not require any additional intervention due to this occurrence.No patient adverse effects have been reported.Reviews of the complaint history, device history record (dhr), instructions for use (ifu), manufacturer¿s instructions (mi), quality control (qc) procedures, as well as a visual inspection and functional testing of the returned devices were conducted during the investigation.Both stents were returned for investigation.There was no way to distinguish which stent was from which lot.Both stents were loaded onto a stock 0.035 wire guide and advanced without difficulty.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record found no non-conformances related to the reported failure mode.A complaint history database search showed no other related complaints associated with the failure mode for the complaint device lot.The information provided upon review of complaint file, device history record, complaint history, and quality control documents did not provide evidence to support that the device was manufactured out of specification or to suggest items in the lot or similar devices in the field or in house were nonconforming.The instructions for use (ifu) provides the following information to the user: precautions ¿ do not force set components during placement, replacement, or removal.Carefully remove the set components if any resistance is encountered.¿ ¿how supplied upon removal from package, inspect the product to ensure no damage has occurred.¿ based upon the available information, inspection of the returned device, and results of the investigation, cook has concluded that the cause for the complaint could not be established.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints per the risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|