Terumo has received the device for evaluation; however, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo references evaluation conclusion code 11.H3: evaluation is in progress, but not yet concluded.
|
|
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during prime, a leakage during priming of the shunt sensor was observed.The shunt sensor was connected to the tube to start priming.After priming was completed, leakage was observed when the pump was stopped.The medical engineer reported that the leakage may have occurred on the small blue luer side at the bottom of the shunt sensor.No patient involvement, product was changed out, procedure was completed successfully.
|
|
This follow-up report is submitted to fda in accord with applicable regulations¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on may 9, 2025.Upon further investigation of the reported event, the following information is new and/or changed: g3 (date received by manufacturer) g6 (indication that this is a follow-up report) h2 (follow-up due to additional information and device evaluation) h3 (device evaluated by manufacturer) h6 (identification of evaluation codes 10, 11, 3331, 4210, 19) the returned sample was visually inspected upon receipt, during which it was noted that the sparger assembly and white luer cap were not present.A white luer cap was added to the sample.It was then leak tested, as received, by connecting with a calibrated manometer, submerged into a water bath, and pressurized up to 1030 mmhg and a leak was noted immediately.The large bore adapter blue cap was then loosened and re-tightened by hand.The sample was then leak tested a second time by connecting with the calibrated manometer, submerged into a water bath, and pressurized up to 1030 mmhg, and no leaks were noted.A retention sample from the same lot number was visually inspected and confirmed to have no traces of buffer on the outside of the unit or inside the pouch and it was then pressurized with air up to 1030 mmhg, submerged in a water bath, and observed for any leaks.No leaks were noted on the retention sample.The root cause for this event was determined to be the large blue vent cap for shunt sensor was not fully tightened either during setup of the circuit, or after the gas calibration.When the large blue vent cap was loosened, it had not been re-tightened fully, causing a leak from the cap.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|